HIOKI SM7110 / SM7120 Super Megohmmeters
• Max. 2000 V output : SM7120
|
Number of channels |
1 ch |
|
DC current measurement |
20 pA range (0.1 fA resolution), Accuracy: ±(2.0 % of rdg. +30 dgt.) 200 pA range (1.0 fA resolution), Accuracy: ±(1.0 % of rdg. +30 dgt.) 2 nA range (10 fA resolution), Accuracy: ±(0.5 % of rdg. +20 dgt.) 20 nA range (100 fA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 nA range (1 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 2 μA range (10 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 20 μA range (100 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 μA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) *2 mA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +30 dgt.) (1) Measurement speed SLOW2 (internal integration time 13PLC) (2) At a temperature of 23 °C ±5 °C with humidity of 85% rh (3) 2 mA range (Measurement speed FAST only) |
|
Resistance measurement capabilities |
50 Ω to 2 × 10 ^19 Ω |
|
Note: Resistance measurement accuracy is defined by the current range accuracy and voltage setting accuracy. |
|
|
Setting voltage range (Accuracy) |
0.1 to 100.0 V, 100 mV resolution, Accuracy: ±0.1 % of setting ±0.05% f.s. 100.1 to 1000 V, 1 V resolution, Accuracy: ±0.1 % of setting ±0.05% f.s. [SM7120 only] 1000 to 2000 V,1 V resolution, Accuracy: ±0.2 % of setting ±0.10% f.s. |
|
Current Limiter |
0.1 to 250.0 V: 5/ 10/ 50 mA, 251 to 1000 V: 5/ 10 mA, to 2000 V:1.8 mA |
|
Measurement time setting |
Delay: 0 to 9,999 ms |
|
Functions |
Comparator, averaging, self-calibration, jig Capacity open correction, cable length correction, surface resistivity, volume resistivity, voltage monitor, contact check |
|
Program function |
10 types of discharge, charge, measure and measurement sequence discharge patterns can be programmed. |
|
Display |
LCD (8 lines of 30 characters), with backlight, High voltage warning indicator |
|
Interfaces |
USB, RS-232C, GP-IB, EXT I/O (NPN/PNP can be switched) |
|
Power supply |
100 to 240V AC , 50/60 Hz, 45 VA |
|
Dimensions and mass |
330 mm (12.99 in)W × 80 mm (3.15 in)H × 450 mm (17.72 in)D, 5.9 kg (208.1 oz) |
|
Accessories |
Power cord ×1, Instruction manual ×1, CD-R (Communications command instruction manual, USB driver) ×1, EXT I/O male connector ×1, Short plug ×1 |
• 300 times better noise resistance
• Max. 2000 V output : SM7120
Max. 1000 V output : SM7110
• Max. 2 × 10^19 Ω display
• Min. 0.1 fA resolution
• Built-in EXT I/O, RS-232C, GP-IB and USB
• Flexible, Multipurpose Design, High Resistance Meter/Electrometer/Picoammeter/IR Meter
• Measure resistance of materials by combining with optional electrode