HIOKI 3276 Current on Probe
• Connects directly to oscilloscope or Memory HiCorder BNC input terminal *1
• High S/N characteristics enable the measurement of 10 mA order current waveforms (3273-50, 3276)
*1 Connecting the device’s metal BNC terminal to a Memory HiCorder’s plastic BNC terminal may damage the plastic terminal or cause it to become misshapen.
|
Frequency bandwidth |
DC to 100 MHz (-3 dB) |
|
Rise time |
3.5 ns or shorter |
|
Noise level |
2.5 mA rms max. (bandwidth limited to 20 MHz) |
|
Continuous allowable input |
30 A rms (requires derating at frequency) |
|
Max. allowable peak input |
50 A peak (non continuous) |
|
Amplitude accuracy (30 min. after power-on, after degaussing and zero-adjustment) |
±1.0 % rdg. ±1 mV f.s. (DC, 45 to 66 Hz, 0 to 30 A rms) ±2 % rdg. (DC, 45 to 66 Hz, 30 A rms to 50 A peak) |
|
Output rate |
0.1 V/A (The output of this probe is internally terminated, so connect to voltage input equipment with high input impedance such as 1MΩ) |
|
Measurable conductors |
Insulated conductor |
|
Core diameter |
φ 5 mm (0.20 in) |
|
Power supply |
±12 V ±0.5 V, 5.3 VA max. |
|
Dimensions and mass |
175 mm (6.89 in)W × 18 mm (0.71 in)H × 40 mm (1.57 in)D, 240 g (8.5 oz) Sensor cable BNC terminal: 1.5 m (4.92 ft), Power cable: 1 m (3.28 ft) |
|
Accessories |
Instruction manual ×1, Carrying case × 1 |
• Waveform observation across a wide band from DC to MHz
• Connects directly to oscilloscope or Memory HiCorder BNC input terminal *1
• High S/N characteristics enable the measurement of 10 mA order current waveforms (3273-50, 3276)
*1 Connecting the device’s metal BNC terminal to a Memory HiCorder’s plastic BNC terminal may damage the plastic terminal or cause it to become misshapen.