• HIOKI CT6710 / CT6711 Current Probe

HIOKI CT6710 / CT6711 Current Probe

• 3 ranges in a single probe - 30 A, 5 A, 0.5 A. Observe a wide current range from micro currents to 30 A.
• Wide band: DC to 50 MHz (-3 dB)
• High S/N ratio and 10 times output rate: Observe waveforms at 100 μA/div at oscilloscope maximum voltage sensitivity setting of 1 mV/div
Model : HIOKI CT6710 / CT6711

Frequency bandwidth

DC to 50 MHz (-3 dB)

Rise time

7.0 ns or shorter

Delay time (Typical)

30 A range: 12 ns, 5 A range: 12 ns, 0.5 A range: 13 ns

(Delay time relative to rising waveform of input signal 1 ns)

Noise level

75 μA rms max (at 0.5 A range, using 20 MHz band measuring instrument)

Max. rated current

30 A range: 30 A rms, 5 A range: 5 A rms, 0.5 A range: 0.5 A rms

(DC, and sine wave, requires derating at frequency)

Max. allowable peak current

30 A range: ±50 A peak (within the input limit time 2 s)

5 A range: ± 7.5 A peak,

0.5 A range: ± 0.75 A peak (< 10 MHz), ±0.3 A peak ( 10 MHz)

Amplitude accuracy

30 A range: ±3.0% rdg. ±1 mV, (Typical) ±1.0% rdg. ±1 mV ( 10 Arms, DC, 45 to 66 Hz sine wave, within the maximum peak current of each range)

5 A range: ±3.0% rdg. ±1 mV, (Typical) ±1.0% rdg. ±1 mV (DC, 45 to 66 Hz sine wave, within the maximum peak current of each range)

0.5 A range: ±3.0% rdg. ±10 mV, (Typical) ±1.0% rdg. ±10 mV (DC, 45 to 66 Hz sine wave, within the maximum peak current of each range)

Output rate

30 A range: 0.1 V/A, 5 A range: 1 V/A, 0.5 A range: 10 V/A, (The output of this probe is internally terminated, so connect to voltage input equipment with high input impedance such as 1MΩ)

Measurable conductors

φ 5 mm (0.20 in), Insulated conductor

Power supply

Supplied from Power Supply 3269, Probe Power Unit Z5021

Cable length

Sensor cable (between relay box and sensor): 1.5 m (4.92 ft)

Power cable: 1.0 m (3.28 ft) (Power plug: FFA.0S.304.CLAC37Y / LEMO inc.)

Dimensions and mass

Sensor: 155 mm (6.10 in)W × 18 mm (0.71 in)H × 26 mm (1.02 in)D, Relay box section: 45 mm (1.77 in)W × 120 mm (4.72 in)H × 25 mm (0.98 in)D

Terminator section: 29 mm (1.14 in)W × 83 mm (3.27 in)H × 40 mm (1.57 in)D mm, 370 g (13.1 oz)

Accessories

Instruction manual ×1, Carrying case ×1

    3 ranges in a single probe - 30 A, 5 A, 0.5 A.  Observe a wide current range from micro currents to 30 A.

    Wide band: DC to 50 MHz (-3 dB)

    High S/N ratio and 10 times output rate: Observe waveforms at 100 μA/div at oscilloscope maximum voltage sensitivity setting of 1 mV/div

    Directly connect to an oscilloscope’s BNC input terminal *1

    Also available: CT6711 Current Probe with DC to 120 MHz bandwidth

*1 Connecting the probe's metal BNC terminal to a Memory HiCorder’s plastic BNC terminal may damage the plastic terminal or cause it to become misshapen. To avoid damage, please position the probe cable perpendicular to the recorder before detaching it from the terminal.

Related Products

HIOKI 3275 Current on Probe