HIOKI 9661 Current on Probe
• Measure up to 500 A
|
Rated measurement current |
500 A AC |
|
Max. measurement current |
Continuous 550 A (45 to 66 Hz) |
|
Output rate |
1 mV AC/ A |
|
Amplitude accuracy (45 to 66 Hz) |
±0.3% rdg. ±0.01% f.s. |
|
Phase accuracy |
±0.5° (45 Hz to 5 kHz) |
|
Amplitude frequency characteristics |
Within ±1% at 40 Hz - 5 kHz (deviation from amplitude accuracy) |
|
Max. rated voltage to earth |
600 V AC rms or less |
|
Measurable conductor diameter |
φ 46 mm (1.81 in) or less |
|
Operating temperature and humidity |
0°C to 50°C (32°F to 122°F), 80% RH or less (no condensation) |
|
Dustproofness and waterproofness |
N/A |
|
Dimensions and mass |
78 mm (3.07 in)W × 152 mm (5.98 in)H × 42 mm (1.65 in)D, 380 g (13.4 oz) Cord length 3 m (9.84 ft), Output terminal: BNC |
• For inverter devices or commercial power lines
• Measure up to 500 A